The ozone chamber is the main factor behind the cracks in rubber material, even if it is rare in the atmosphere. This machine simulates and strengthens the ozone condition. It can study the effect of it on rubber and find solutions to improve product life.
Ozone Test Chamber Specification
Model | XB-OTS-80S | XB-OTS-150S | XB-OTS-225S | XB-OTS-408S | XB-OTS-800S | XB-OTS-1000S |
Inner chamber size W*H*D(cm) | 40*50*40 | 50*60*50 | 50*75*60 | 60*85*80 | 100*100*80 | 100*100*100 |
Exterior chamber size W*H*D(cm) | 90*90*150 | 95*150*105 | 95*165*115 | 105*175*135 | 145*190*135 | 145*190*155 |
Temperature Range | RT to 90°C | |||||
Humidity Range | 45%RH ~98%RH | |||||
Temp.&Humidity Fluctuation | ±0.2°C, ±0.5%RH | |||||
Temp.&Humidity Uniformity | ±1.5°C,±2.5% RH when RH is ≤ 75%,±4% RH when RH is > 75% (tested with the empty chamber, 30 minutes after stabilization) | |||||
Temp.& Humidity Resolution | 0.01°C,0.1%RH | |||||
Ozone Concentration | 0 to 1000 PPHM | |||||
Ozone Control Accuracy | ±10% | |||||
Ozone Generation | Static emissions | |||||
Sample Rotating Speed | 1 times per minute | |||||
Exterior Chamber Material | Stainless steel plate+Powder Coated | |||||
Interior Chamber Material | SUS#304 stainless steel plate | |||||
Insulation | PU & Fiberglass wool | |||||
Ozone Analyzer | Imported ozone density analyzer | |||||
Ozone Generator | Silent discharge type ozone generator | |||||
Wind Circulation | Sirocco fan | |||||
Heating System | SUS#304 stainless steel high-speed heater | |||||
Humidifying | Surface evaporating system | |||||
Cooling | Imported compressor,finned type evaporator,air-cooling condenser | |||||
Dehumidifying | ADP critical dew-point cooling/dehumidifying method | |||||
Controller | Digital electronic indicators+S.S.R.With PID automatic calculation capability | |||||
Other Components | Vacuum glass inspection window,cable port(50mm),control status indicator lights, chamber lamp,loading shelves (2pcs for free) | |||||
Safety Devices | Overheating circuit-breaker,compressor overload protection,control system overload protection, humidifying system overheating protection, breakdown indicator light | |||||
Power Supply | AC 1 phase 220V; 3 phase 380V 60/50Hz |
Ozone Chamber Features
-
External chamber: Arc design, electrostatically sprayed coat.
-
Inner chamber: SUS#304 stainless plate, anti-corrosion, and water resistance.
-
The door has a transparent window (220*320mm) for observation (lighting inside). It has the advantages of transparency, heat insulation, and explosion-proof.
-
The air duct of the windway is in the same direction as the specimen growth. A high-tensile silicone seal is used between the test door and body to ensure the test area is closed.
-
Cooling system: Installed at the bottom. The Tecumseh compressor is originally from France.
-
High-quality fixed wheels in the bottom, easy to move.
-
Heating system: completely independent, nickel-chromium alloy electric heater.
-
Heater control way: contact cycle pulse width modulated, SSR(solid state relays)
-
Output power calculus by microcomputer to control a smooth, reliable.